Alarms
6-8
6.3.5
Status Switchover
1. In the Normal status,
Press the SILENCE key for less than 2 seconds to switch to the Alarms Paused
status, or
Press the SILENCE key for 2 seconds or more to switch to the System Silenced
status.
2. In the Alarms Paused status,
Press the SILENCE key for less than 2 seconds to switch to the Normal status,
or
Press the SILENCE key for 2 seconds or more to switch to the System Silenced
status.
3. In the System Silenced status,
Press the SILENCE key for less than 2 seconds to switch to the Alarms Paused
status, or
Press the SILENCE key for 2 seconds or more to switch to the Normal status.
4. In the Alarms Silenced status,
Enter the ALARM SETUP menu and set ALARM VOL to any setting other
than 0. The alarm sound is restored.
Enter the USER MAINTAIN menu and set ALARM SOUND OFF to
DISABLE. The system restores the default alarm volume.