SpO
2
Monitoring
12-10
ALM
SpO
2
alarm on/off status
ON: When a SpO
2
alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a SpO
2
alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right
of the SPO2 label.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
ALM REC
Alarm recording
ON: When a SpO
2
alarm occurs, the monitor enables the
recording;
OFF: When a SpO
2
alarm occurs, the monitor does not
enable the recording.
SPO2 ALM HI
SpO
2
upper alarm limit
SPO2 ALM LO
SpO
2
lower alarm limit
PR ALM HI
PR upper alarm limit
PR ALM LO
PR lower alarm limit
SpO
2
and PR alarm limits:
Parameter
Max. upper limit
Min. lower limit
Step
SpO
2
100 0
1
PR 254 0 1
The default SpO
2
and PR alarm limits:
Parameter Patient
type Upper limit
Lower limit
Adult 100 90
Pediatric 100
90
SpO
2
Neonate 95
80
Adult 120 50
Pediatric 160
75
PR
Neonate 200 100
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...