Appendices
20-36
C.2.13
Welch Allyn CO
2
Module Alarm Messages
Alarm message
A
B
L
Cause
Measure
CO2 SENSOR OFF
No
Yes
3
The mainstream
sensor is not
properly
connected or is
disconnected.
Make sure the sensor is properly
connected.
CO2 SIGNAL LOW
No
No
3
The CO
2
signal is
weak.
CO2 SIGNAL TOO LOW
No
No
3
The CO
2
signal is
too weak.
Check for leaks in the airway.
Check if the airway is clogged.
Check if the water trap is too
old. If all the above problems are
eliminated but the problem still
exists, replace with a new CO
2
tube or water trap. If the problem
remains, contact our company
for repair.
CO2 BAROMETRIC TOO
LARGE
No No 2
CO2 PNEUMATIC LEAK
No
No
2
CO2 SIGNAL NOISY
No
No
3
CO2 SIGNAL SATURATE
No
No
3
CO2 CALCULATION ERR
No
No
1
CO2 SENSOR FAULT
No
No
1
CO2 SENSOR TEMP HIGH
No
No
1
CO
2
module fault
Contact our company for repair.
CO2 SENSOR TEMP LOW
No
No
1
CO2 WATCHDOG TIMEOUT Yes
No
1
CO2 SYSTEM ROM ERR 1
Yes
No
1
CO2 FLASH CRC ERR
Yes
No
1
CO2 INT COMM ERR
Yes
No
1
CO2 EXT RAM ERR
Yes
No
1
CO2 INT RAM ERR
Yes
No
1
CO2 FLASH CHECK ERR
Yes
No
1
CO2 STACK OVER
Yes
No
1
CO2 PUMP FAULT
No
No
1
CO2 REVERSE FLOW
No
No
1
CO2 FORWARD FLOW
No
No
1
CO2 MALFUNCTION
No
No
1
CO2 BAROMETRIC HIGH
No
No
1
CO2 BAROMETRIC LOW
No
No
1
CO
2
module fault
Contact our company for repair.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...