ECG/RESP Monitoring
11-16
NOTE
z
The SMART LEAD OFF function is available only when the LEAD TYPE is
5 LEADS and 12 LEADS.
z
When the HR CHANNEL is set to AUTO, SMART LEAD OFF does not
work as long as there is ECG waveform displayed in either channel 1 or
2.
z
When you perform ECG calibration, SMART LEAD OFF does not work.
BEAT VOL
Beat volume
Range : 0-10. 0 indicates disabled and 10 indicates
the maximum volume.
PACE
ON: When ON is selected, a detected pacemaker
signal is indicated by a "
︱
" symbol above the ECG
waveform.
OFF: When OFF is selected, the pacemaker
analysis is disabled.
PACE DETECT
Pace detection
Options: MODE1 and MODE2
1. MODE1: In this mode, the function of pacemaker pulse rejection with
overshoot is enabled.
2. MODE2: In this mode, the function of pacemaker pulse rejection with
overshoot is disabled.
A pacemaker may produce an overshoot signal after the pacemaker pulse, which
probably causes false R-wave detection. For this reason, the function of pacemaker
pulse rejection with overshoot should be enabled in the calculation. If this function
is enabled for a pacemaker without overshoot signals, the R-wave detection might
be omitted. Hence, the doctor should choose the PACE DETECT mode carefully
according to the pacemaker type.
The PACE DETECT option is active only if PACE is turned ON. When PACE is
turned OFF, the PACE DETECT is switched to MODE1 automatically. In this
situation, you cannot adjust the PACE DETECT as desired. When PACE is turned
ON, you can choose from the PACE DETECT options.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...