Appendices
20-22
Appendix B EMC
The equipment meets the requirements of IEC 60601-1-2:2001.
NOTE
z
Use of accessories, transducers, and cables other than those specified
may result in increased emission and/or decreased immunity of the
equipment.
z
The equipment should not be used adjacent to or stacked with other
equipment, and if adjacent or tacked use is necessary, the equipment
should be observed to verify normal operation in the configuration in
which it will be used.
z
The equipment needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information
provided below.
z
The equipment may be interfered with by other equipment, even if that
other equipment complies with CISPR emission requirements.
z
Operation of the device, in the case that the patient physiological signal
is lower than the minimum amplitude and/or value specified in the
product specifications, may cause inaccurate results.