IBP Monitoring
15-6
OFF: When a IBP alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right
of the IBP (1,2) label.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
AMP ADJUST
Amplitude adjustment
This option allows you to adjust the waveform amplitude.
Options : MANUAL and AUTO.
1. MANUAL: If MANUAL is selected, you can select an IBP channel manually
from the options, including ART, PA, CVP, RAP, LAP, ICP, P1 and P2, and
then adjust the IBP scale in the SCALE ADJUST >> menu.
2. AUTO: If AUTO is selected, the waveform labels of IBP channels become P1
and P2 and the IBP scale is adjusted by the system automatically.
SWEEP
Waveform speed
Options: 12.5 and 25.0 mm/s.
UNIT
Options: mmHg and kPa.
FILTER
This option allows you to select the filter method of the IBP
waveform. Three options are available:
NORMAL: 16Hz filtering;
SMOOTH: 8Hz filtering;
NO FILTER: Displays the original waveform.
The default is NO FILTER.
ALM LIMIT
SETUP >>
This option allows you to access the following menu,
whereby you can set the upper and lower alarm limits.
Figure 15-6 IBP Alarm Setup Menu
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...