Recording
8-3
menu. You can also select the two waveforms (REC WAVE 1 and REC WAVE 2) to
be recorded. If one of the two waveforms is set to OFF, the recorder will merely
print out one waveform and all the measured parameters; if both are set to OFF, the
recorder will only print all the measured parameter.
Automatic recording
The monitor starts recording at the selected interval (TIMING REC TIME) and
record for 8s waveforms automatically. For details, see
4.4.4 Recorder Setup
.
Alarm recording
Alarm recording includes measured parameter recording, ST segment alarm
recording and arrhythmia alarm recording.
1. Parameter alarm recording
When a parameter alarm occurs, the recorder automatically records two waveforms
of 8, 16 or 32 seconds (respectively 4, 8 or 16 seconds before and after the alarm.
See
4.4.2 Alarm Setup
) and all measured parameters.
2. ST segment alarm recording
In case of an ST segment alarm, the recorder automatically prints two ECG
waveforms of 8, 16 or 32 seconds (respectively 4, 8 or 16 seconds before and after
the alarm) and all the measured parameters.
3. Arrhythmia alarm recording
In case of an arrhythmia alarm, the recorder automatically prints an ECG waveform
of 8 seconds (respectively 4 seconds before and after the alarm) and all the measured
parameters.
NOTE
z
For a parameter alarm recording, you must first set the ALM and ALM
REC options to ON.
Frozen waveform recording
In the frozen mode, the monitor can print the frozen waveforms displayed on the
screen and the parameters measured at the moment when the waveforms were frozen.
For details, see
7.5 Recording Frozen Waveforms
.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...