Anesthesia Gas Monitoring
18-6
18.3 AG Setup Menu
Select the GAS label in the parameter window. The AG SETUP menu appears.
Figure 18-4 AG SETUP menu
In this menu, you can set the following items.
Agent
Used to select the name of the anesthetic to be monitored,
including AA, HAL, ENF, ISO, SEV and DES.
If you select the AION 02 AG module, the patient monitor cannot identify the
anesthetic type. Therefore, you have to select the used anesthetic before using the
AG module so that the AG module can conduct measurements normally.
If you select the AION 03 AG module, the used anesthetic will be automatically
displayed here, so you need not set this item.
CO2 UNIT
Options: mmHg, kPa and %.
O2 UNIT
Options: mmHg, kPa and %.
N2O UNIT
%
AA UNIT
%
PUMP RATE
Used to select the appropriate pump rate
3 options: HIGH, MED, LOW
O2 COMPEN
The options include: OFF, 30%, 40%, 50%, 60%, 70%,
80%, 90%, and 100%.
When the O
2
concentration exceeds 60% and it is not being monitored, turn on this
switch.