ECG/RESP Monitoring
11-46
For different patient types, the upper/lower limits of the respiration rate alarm may
vary in the following range.
Patient type
Max. ALM HI
Min. ALM LO
Increment
Adult 120
0
1
Neonate/pediatric 150
0
1
APNEA ALM
Apnea alarm
Determines whether the patient’s cessation of breath is an
apnea event. Range: 10
–
40 seconds.
SWEEP
Waveform speed
Options: 6.25, 12.5 and 25.0 mm/s.
HOLD TYPE
Calculation type
Options: AUTO and MANUAL
1. AUTO: When AUTO is selected, the monitor automatically determines the
detection threshold for respiration and calculates the respiration rate. The
HOLD HI and HOLD LO options are inactive.
2. MANUAL: When MANUAL is selected, the user determines the detection
threshold for respiration and the monitor calculates the respiration rate
depending on the user-selected criteria.
DEFAULT >>
Select DEFAULT >> to access the RESP DEFAULT
CONFIG menu. You can select either FACTORY
DEFAULT CONF or USER DEFAULT CONF. After
finishing the selection and exiting the menu, a dialog pops up
asking for confirmation of your selection.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...