ECG/RESP Monitoring
11-26
11.5.2 Arrhythmia Analysis Menu
Select ARR ANALYSIS >> in ECG SETUP menu. The following menu appears.
Figure 11-13
Arrhythmia Analysis
In this menu, you can perform the following settings:
ARR ANAL
Arrhythmia analysis
ON: Enables the arrhythmia analysis;
OFF: Disables the arrhythmia analysis.
PVCs ALM
PVCs alarm
ON: When the PVCs alarm occurs, the monitor gives alarm
indications and saves the alarm;
OFF: When the PVCs alarm occurs, the monitor neither gives
alarm indications nor saves the alarm;
When OFF is selected, the
icon is displayed on the right of
PVCs in the parameter window.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
ALM REC
Alarm recording
ON: The monitor starts recording when the PVCs alarm occurs.
OFF: The monitor does not record when the PVCs alarm
occurs.
ALM HI
Upper alarm limit
Determines the upper limit of the PVCs alarm. Range: 1
–
10.
An alarm is triggered when the PVCs exceeds the upper limit.
ARR
RELEARN
Arrhythmia relearning
You can select this option to start a learning procedure. The
“ARR LEARNING” message is displayed in the information
area of the screen.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...