ECG/RESP Monitoring
11-2
11.1 Overview
11.1.1 ECG Waveform
In the standard screen, one or two ECG waveform(s) is (are) displayed at the top of
the display when LEAD TYPE is set to 3 LEADS or 5LEADS respectively in the
ECG SETUP menu.
Figure 11-1 ECG Waveforms
As shown above, when 5 LEADS is selected, five labels are located above the ECG
waveforms:
1. ECG lead of channel 1
You can select the lead of channel 1 from the label options.
When you use a 3-lead set, the monitor gives three lead options: I, II and III;
When you use a 5-lead set, the monitor gives seven lead options: II, III, aVR,
aVL, aVF, V and I.
2. Gain of the channel-1 waveform
You may use this label to adjust the amplitude of the channel-1 ECG waveform. The
gain options include ×0.125, ×0.25, ×0.5, ×1, ×2 and AUTO. When AUTO is
selected, the monitor adjusts the gain automatically. Besides, a 1mV scale is
displayed at the right side of each ECG waveform. The height of the 1mV bar is
directly proportional to the ECG waveform amplitude.
3. Filter
Method
The filtering enables clearer and more detailed waveforms. There are three filter
methods for selection.
1 2
3
4 5
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...