NIBP Monitoring
13-9
DEFAULT >>
You can select this option to access the NIBP DEFAULT
CONFIG menu, in which you may select FACTORY
DEFAULT CONFIG or USER DEFAULT CONFIG. After
finishing your selection and exiting the menu, the system
pops up a dialog box asking for your confirmation.
13.4.1 Calibration
If you select the CALIBRATE option, the monitor starts the NIBP calibration and
the CALIBRATE option changes to STOP CAL. Selecting the option again stops the
calibration.
Calibrate the cuff pressure reading with a calibrated reference manometer (or
mercury manometer) with accuracy higher than 1mmHg. To perform the calibration,
follow the procedure shown below:
1. Remove the blood pressure cuff from the monitor and replace it with a rigid
metal container or vessel with a capacity of 500 ml ± 5%.
2. Connect a calibrated reference manometer (with an error less than 1mmHg) and
a ball pump using “T” connectors as shown below.
3. Select the CALIBRATE option.
4. Inflate the metal container using the ball pump until the reference manometer
reads 0, then 50, and finally 200 mmHg.
5. The difference between the indicated pressure of the reference manometer and
the indicated pressure of the monitor will not exceed 3 mmHg. Contact our
Customer Service if these values are not met.
Figure 13-3 NIBP Calibration
Monitor
NIBP cuff
connector
Hose
Reference
manometer
Metal vessle
Ball pump
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...