Appendices
20-18
A.15.2 Oridion
CO
2
Specifications
CO
2
measurement range
0 to 99mmHg
Precision*
0 to 38 mmHg:
39 to 99 mmHg:
±2mmHg
±5% + 0.08%× (reading - 38mmHg)
Resolution
Waveform:
Value:
0.1mmHg
1mmHg
Flow rate
7.5
15
50
−
+
ml/min
Initialization time
30s (typical)
Response time
2.9s (typical)
Delay time
2.7s (typical)
AwRR measurement range 0 to 150 BrPM
AwRR measurement
precision
0 to 70BrPM:
70 to 120BrPM:
121 to 150BrPM:
±1BrPM
±2BrPM
±3BrPM
Apnea alarm delay
AwRR: 10 to 40s
* Precision applies for breath rates of up to 80 bpm. For breath rates above 80 bpm, accuracy
complies with EN 864/ISO 9918 (4 mmHg or ±12% of reading whichever is greater) for EtCO
2
values exceeding 18 mmHg. To achieve the specified accuracies for breath rates above 60
breaths/minute, the Microstream® FilterLine H Set for Infant/Neonatal (p/n 006324) must be
used.
The accuracy specification is maintained to within 4% of the values indicated in the above table
in the presence of interfering gases according to EN864 Section Eleven, Part 101.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...