Appendices
20-44
NIBP Groups
Multiple NIBP groups are
displayed in the lower left
corner of the screen.
Select the NIBP label and press the control knob. In
the NIBP SETUP menu, set DISPLAY WAY to
GROUPS.
Alarm Sound Off
Used to disable the alarm
sound of the patient monitor.
Press the MENU button. In SYSTEM MENU, select
MAINTAIN >>. Enter the user key and confirm it to
enter the USER MAINTAIN menu. Then set
ALARM SOUND OFF to ENABLE.
Pace maker pulse
rejection
The monitor rejects pacemaker
pulses that have amplitudes
from -700 mv to +700 mv,
pulse widths less than 2 ms
and no overshoot
None
Nurse Call
The monitor alerts nurses
when certain alarms are
present.
Press the MENU button. In SYSTEM MENU, select
MAINTAIN >>. Enter the user password and
confirm it to enter the USER MAINTAIN menu. Set
AUX OUTPUT to NURSE CALL, and then select
NURSE CALL SETUP >>. In the NURSE CALL
SETUP menu, set the parameters.
Data Output
The data can be output from
the patient monitor to the PC.
Press the MENU button. In SYSTEM MENU, select
SYSTEM SETUP >> to enter the SYSTEM SETUP
menu, and then select DATA OUTPUT >>. In the
DATA OUTPUT sub-menu, set parameters, and
then select the OUTPUT button.
Large Font Screen
The monitored parameters and
information can be displayed
in large fonts.
Press the MENU button. In SYSTEM MENU, select
the SYSTEM SETUP >> button. In the SYSTEM
SETUP menu, select LARGE FONT SCREEN in
the FACE SELECT sub-menu.
Defibrillator
Synchronization Pulse
Signal
In case of ventricular
fibrillation, the defibrillator
synchronization pulse signal
(100ms, +5V) can be output to
the defibrillator through the
auxiliary output interface on
the back of the patient
monitor.
Press the MENU button. In SYSTEM MENU, select
MAINTAIN >>. Enter the user key and confirm it to
enter the USER MAINTAIN menu. Then set AUX
OUTPUT to DEFIB. SYN. Then
Select the ECG label and press the control knob. In
the ECG SETUP menu, select OTHER SETUP >>,
and then set DEFIB SYNC to ON.
Non-volatile
Measurement data can be kept
in the monitor memory even
after the monitor is powered
off.
None
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...