Freezing Waveforms
7-3
Press the FREEZE key on the control panel again.
After exiting the frozen mode, all waveforms on the screen are cleared and new
real-time waveforms are displayed. If the scan type of the monitor is set to
REFRESH, the waveforms are refreshed from the left of the waveforms area to the
right ; if the scan type is set to SCROLL, the waveforms are displayed from the right
of the waveforms area to the left and are scrolling.
7.3
FROZEN Menu
The FROZEN menu is displayed at the lower left corner. You can perform the
following settings in this menu.
WAVE 1 Waveform 1
It determines the first frozen waveform to be printed on the recorder
paper. The options of waveform 1 include all waveforms displayed
on the screen and OFF. The waveform 1 will not be recorded when
OFF is selected.
WAVE 2 Waveform 2
It determines the second frozen waveform to be printed on the
recorder paper. The options of waveform 2 include all waveforms
displayed on the screen and OFF. The waveform 2 will not be
recorded when OFF is selected.
RECALL Waveform recall
It changes to L-RIGHT when selected. At this time, you can rotate
the control knob to view the frozen waveform.
REC
Record
When selected, the recorder starts printing out the selected
waveform 1 and waveform 2.
EXIT
This option allows you to exit the FROZEN menu and unfreeze the
waveforms.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...