Alarms
6-6
6.3
Alarm Statuses
When an alarm occurs, normally the monitor gives indications in the modes
mentioned above as per the alarm level. If necessary, you can the set the monitor to
the following alarm statuses.
Alarms Disabled
Alarms Paused
System Silenced
Alarms Silenced
6.3.1
Alarms Disabled
If the alarm switch of a parameter is turned off, the monitor does not generate alarms
even if the measured parameter value exceeds the alarm limit. This status is called
Alarms Disabled.
To disable the alarms of a parameter, you should open the setup menu of the
parameter. Take heart rate (HR) as an example.
1. Rotate the control knob and highlight the ECG parameter label.
2. Press the control knob. The ECG SETUP menu pops up.
3. Rotate the control knob and highlight the field on the right of HR ALM.
4. Press the control knob, and then select OFF from the pull-down list.
5. The HR alarms are disabled. The
icon will be displayed on the right of the
ECG parameter label.
NOTE
z
When a new parameter module is installed or when a parameter module
is turned ON, all the parameter alarms and technical alarms related to
this module are disabled in the first 30-second operating time. The other
module alarms are not influenced.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...