ECG/RESP Monitoring
11-39
ST Analysis
Select the ST label in the parameter area, or select ST ANALYSIS >> in ECG
SETUP menu. The following menu appears.
Figure 11-27 ST ANALYSIS menu
In the ST ANALYSIS menu, you can perform the following settings:
ST ANAL
ST analysis
ON: Enables the ST analysis;
OFF: Disables the ST analysis.
NOTE
z
When ST ANAL is set to ON, the monitor selects DIAGNOSTIC mode
automatically. You can set the monitor to MONITOR or SURGERY mode
as required. However, the ST numerics might be severely distorted in
these modes.
ST ALM
ST segment alarm
ON: If the measured ST numerics exceed the alarm limit, the
monitor gives alarm indications and saves the alarm;
OFF: If the measured ST numerics exceed the alarm limit, the
monitor does not give alarm indications or save the alarm.
When OFF is selected, the
icon is displayed on the upper
right of the ST data area.
ALM LEV
ST alarm level
Options: HIGH, MED and LOW;