IBP Monitoring
15-11
15.4.3 IBP Pressure Calibration
You can select IBP PRESSURE CALIBRATE in IBP (1,2) SELECT menu to open
the following menu.
Figure 15-11 IBP Pressure Calibration Menu
Calibration Procedure
1. Disconnect the pressure transducer from the patient. Connect the 3-way
stopcock, the sphygmomanometer and the inflation orb, using a T-shape
connector, as shown below.
Figure 15-12 IBP Pressure Calibration
3-way
stopcock
Pressure
transducer
Pressure transducer
interface cable
Monitor
Hydrargyrum
pressure meter
T-shape connector