IBP Monitoring
15-23
15.6.5 Maintenance and Cleaning
WARNING
z
The ICT/B is supplied non-sterile. It must be cleaned and sterilized each
time before use.
z
Do not autoclave; do not use radiation sterilization or ultrasonic
cleaning.
z
Do not use the sterilizing cap during ETO gas sterilizing.
z
Do not use chlorinated hydrocarbons, toluene, or sodium hypochlorite
solution for disinfection.
Right after the catheter is removed from the patient, check the silicone coating for
damage. Use a 1ml syringe to inject 0.5cc air into the luer and immerse the catheter
in water. If small bubbles are seen from any part of the catheter or tip, dry the
catheter by a cloth and sterilize. Then, return to the manufacturer for repair.
Inspect for cuts or damage to silicone coating before immersing it in any liquid.
Avoid getting liquid on the connector pins or inside the connector via the luer fitting.
Wash the catheter with soap solution. Avoid damaging the sensing area. Do not use
synthetic detergents or oil-based soaps as this may result in a foreign body reaction.
Transducers may be cleaned gently with alcohol. But, do not soak it in alcohol. The
cold aqueous solutions of detergicide (e.g. Cidex), formalin or by ethylene oxide gas
can be used for disinfection.
NOTE
z
It is recommended that each institution establish a sterilization
procedure to specify what products are to be disinfected.
z
There are only two chemical sterilization techniques recognized by the
U.S. Department of Agriculture as effective and truly sporicidal, gas
sterilization by ethylene oxide and liquid sterilization by a
glutaraldehyde.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...