CO
2
Monitoring
17-11
ZERO CAL
This option allows you to zero the CO
2
module, so as to
eliminate the adverse affect of the baseline drift in the
process of measurement. Perform zeroing prior to
concentration calibration.
CONFIRM CAL This option allows you to confirm the calibration of the CO
2
module.
Calibration Procedure
1. Verify the module enters the ready-to-measure status.
2. Connect a gas bottle with the CO
2
sensor connector on the monitor as shown
below.
Figure 17-6 Connections for Calibration
3. Fill the gas bottle with a standard gas of certain CO
2
content (3%, 4%, 5%, 6%
and 7%), and input the gas to the monitor.
4. Open the CO2 USER MAINTAIN menu, and set the CO
2
field to a value,
which is the same with the CO
2
content of the gas bottle.
5. When the CO2 USER MAINTAIN menu displays the currently measured CO
2
content, wait until a stable reading is indicated, and then select the CONFIRM
CAL option to start calibrating the CO
2
module.
6. If the calibration succeeds, “CALIBRATE SUCCESS!” will be displayed in the
CO2 USER MAINTAIN menu. If fails, “CAL FAIL, TRY AGAIN!” will be
displayed, and re-calibration is required.
7. Select the EXIT option to exit the CO2 USER MAINTAIN menu.
Monitor
Gas bottle
T-shape connector
opening to the
atmosphere
Decompression
valve
Tube
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...