ECG/RESP Monitoring
11-10
11.2.2.3 Electrode Placement for Surgical Patients
Electrode placement during surgery is dependent on the type of surgery being
performed. For example, with open chest surgery, the electrodes might be placed
laterally on the chest or on the back. In the operating room, artifact can sometimes
affect the ECG waveform due to the use of electrosurgery equipment. To help reduce
this, place the electrodes on the right and left shoulders, the right and left sides near
the stomach, and place the chest lead on the left side at mid-chest. Avoid placing the
electrodes on the upper arms. This will cause the ECG waveform to be too small.
WARNING
z
When using Electrosurgery equipment, patient leads should be placed in
a position that is equal distance from the Electrosurgery electrotome
and the grounding plate to avoid burns to the patient. The
Electrosurgery equipment wire and the ECG cable must be kept
separated and not allowed to tangle.
z
When using Electrosurgery equipment, never place the ECG electrodes
near the grounding plate of the Electrosurgery device. This will cause a
great deal of interference with the ECG signal.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...