Appendices
20-30
Alarm message
A
B
L
Cause
Measure
SIGNAL
SATUATED
No Yes 3
RANGE
EXCEEDED
No Yes 3
or calculation.
EXCESSIVE
MOTION
No
Yes 3
Excessive motion of the patient’s
arms.
OVER PRESSURE
No
Yes 2
The airway might be blocked.
NIBP SYSTEM
FAILURE
No Yes 2
NIBP TIME OUT
No
Yes 2
MEASURE FAIL
No
Yes 2
Failure occurs in the pulse
measurement. The monitor cannot
perform measurement, analysis,
or calculation.
Check the patient’s condition,
and check if the patient type is
correct. Replace with a proper
cuff and connect it correctly. If
the problem still exists, contact
our company for repair.
NIBP ILLEGALLY
RESET
No
Yes 2
Illegal reset comes out during the
NIBP measurement.
Check if the airway is blocked.
Deal with the blocking and
perform the measurement again.
If the problem still exists,
contact our company for repair.
C.2.6 Mindray
SpO
2
Module Alarm Messages
Alarm message
A
B
L
Cause
Measure
SPO2 SENSOR
OFF
No
Yes 3
The sensor is disconnected from
the patient or the monitor.
Make sure the sensor is placed
on the patient’s finger or other
parts, and the monitor is
connected to cables correctly.
The sensor is disconnected from
the patient or the monitor, or it is
not properly connected.
Disconnect and reconnect the
sensor as directed by the
instructions. If the alarm
remains, the sensor or the cable
might have been damaged.
SPO2 NO SENSOR
Yes Yes 3
The SpO
2
is connected upside
down.
Disconnect and reconnect the
sensor as directed by the
instructions. Pay attention to the
mark on the probe.
SPO2 LOW
PERFUSION
No
No
3
The pulse signal is too weak.
Move the sensor to a site with
better perfusion.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...