ECG/RESP Monitoring
11-27
11.5.3 Arrhythmia Alarm Setup
Select ARR ALARM >> in ECG SETUP menu. The following menu appears. You
can change the settings of the arrhythmia alarm in this menu.
Figure 11-14
Arrhythmia Alarm Setup
In the menu, the ALM field indicates the alarm on/off status, REC indicates the
alarm recording on/off status and LEV indicates the alarm level. You can change the
settings as described below.
ALL ALM ON
All alarms on
Enables all arrhythmia alarms;
ALL ALM OFF
All alarms off
Disables all arrhythmia alarms;
ALL REC ON
All recording on
Enables the recording of all arrhythmia alarms;
ALL REC OFF
All recording off
Disables the recording of all arrhythmia alarms;
ALM LEV
Options: HIGH, MED and LOW.
Sets the level of all arrhythmia alarms to the same value.