Anesthesia Gas Monitoring
18-2
18.1 Overview
The anesthesia gas (AG) monitoring can be used for measuring the anesthesia gas
and respiration gas of the patient in the anesthetic status. This monitor can configure
AION 02 AG module or AION 03 AG module. AG module provides the end-tidal
numerics and inhaling numerics of the gases mentioned below.
Carbon dioxide (CO
2
): The measured numeric is EtCO
2
(Max. exhaling value:
Max. exhaling numeric detected during the respiration).
Nitrous oxide (N
2
O): Laughing gas.
Oxygen (O
2
): Optional function.
Anesthetic (AA): Refers to the monitored anesthetic (DES, ISO, ENF, SEV or
HAL).
Airway respiration rate (AwRR): respiration per minute (BrPM).
The patient monitor can display simultaneously a maximum of 4 waveforms,
including the CO
2
waveform (default waveform), N
2
O waveform, O
2
waveform and
the anesthetic (ENF: Enflurane) waveform.
Figure 18-1 AG waveform
In addition, the patient monitor can display parameters, including CO
2
, N
2
O, O
2
and
AA (Which refers to the monitored anesthetic: DES, ISO, ENF, SEV or HAL). It
also displays the inhaling and exhaling numerics as well as MAC (Minimum
Alveolar Concentration)/MAL (balance gas) and AwRR.
When the O
2
module is not connected, no O
2
waveform will be displayed.
After the O
2
module is connected, whether the O
2
waveform is displayed depends on
Waveform
name
Waveform
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...