Appendices
20-28
C.2
Technical Alarm Messages
Note: XX represents the parameter modules like ECG, NIBP and SpO
2
, or the
parameters being monitored like HR, PR and SpO
2
.
The A field indicates whether an alarm can be completely cleared; the B field
indicates whether the visual and audible indications of an alarm can be cleared; the
“L” field indicates the alarm level, and 1 means high, 2 means medium, 3 mean low,
* means the level is user-adjustable.
C.2.1
General Alarm Messages of Parameter Modules
Alarm
message A B L Cause
Measure
XX INIT ERR N
Yes No
1
XX module initialization error N
Note: N stands for the error number.
XX COMM STOP
No
No
1
Failure in communication between XX
module and the main board.
XX COMM ERR
Yes No
1
Error in communication between XX
module and the main board.
Restart the monitor.
If the error remains,
contact our
company for repair.
XX ALM LMT
ERR
No
No
1
The alarm limit of the XX parameter is
changed inadvertently.
XX EXCEED
No
No
1
The measured XX parameter value exceeds
the measurement range.
If the error remains,
contact our
company for repair.
C.2.2
ECG Module Alarm Messages
Alarm
message A B L Cause
Measure
ECG LEAD OFF
No
Yes 3
ECG X LEAD OFF
No
Yes 3
The ECG lead is not connected
correctly.
Note: X represents the leadwires, V, LL, RL, LA and RA, as per AHA standard,
or C, L, F, R and N as per IEC standard.
Check for correct connection
of the leadwires.
ECG NOISE
No
No
3
Large interference signals appear
on the ECG signal.
Make sure the leadwires are
correctly connected. Check
the patient for severe
motion.
ECG CH1
SELFTEST ERR
Yes No
1
An error occurs in the ECG
initialization.
Restart the monitor. If the
error remains, contact our
company for repair.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...