ECG/RESP Monitoring
11-15
ECG DISPLAY ECG display mode
Three options are available:
1. NORMAL DISPLAY: The monitor displays two ECG waveforms when the
5-lead set is used or one ECG waveform when the 3-lead set is used.
2. MULTI-LEADS
DISPLAY:
The monitor displays six ECG waveforms, which
occupies the whole waveforms area.
3. HALF-SCN MULTI-LEADS: The monitor displays six ECG waveforms,
which only occupies four waveform positions.
NOTE
z
If 3 LEADS is selected from the LEAD TYPE options, only NORMAL
DISPLAY can be selected for the ECG DISPLAY.
SMART LEAD
OFF
ON: If there is a LEAD OFF in the HR-derived channel,
the system will automatically switch to another lead to
restore the display of the ECG waveform in the
HR-derived channel.
OFF: The SMART LEAD OFF is turned off.
When ECG is in NORMAL DISPLAY mode, if the SMART LEAD OFF is
configured ON and there is a LEAD OFF in the HR-derived channel, the system will
automatically switch to another lead to restore the display of the ECG waveform in
the HR-derived channel. When the LEAD OFF condition is corrected, the leads are
automatically switched back. However, if you manually switch the HR-derived
channel, change the ECG display mode, or turn off SMART LEAD OFF before the
LEAD OFF condition is corrected, the leas will not be switched back.
Switching rules for SMART LEAD OFF:
LEAD OFF condition
HR-derived channel automatically switches to
RA OFF
III
LL OFF
I
LA OFF
II
V OFF
II
RL OFF
None
Any two out of LL, RA and LA fall off.
None
LL, RA or LA, and V fall off together
Same as LL, RA or LA falls off.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...