NIBP Monitoring
13-11
NOTE
z
Set PAT TYPE to ADU in the PATIENT SETUP menu before leakage test.
z
The pneumatic test, other than being specified in the EN 1060-1
standard, is to be used to simply determine whether there are air leaks in
the NIBP airway.
13.5 Maintenance and Cleaning
WARNING
z
Do not press the rubber tube on the cuff with excessive strength.
z
When a reusable cuff is not connected to the monitor or is being
cleaned, always cover the rubber tube with a lid. Avoid splashing liquid
into the rubber tube or the monitor inadvertently.
Reusable Blood Pressure Cuffs
The cuff is not suitable for dry-cleaned. Instead, it should be machine or hand
washed. Hand washing, may prolong the service life of the cuff. Before washing,
remove the latex rubber bladder. Allow the cuff to dry thoroughly after washing, and
then reinsert the rubber bladder. The cuff can be disinfected by means of
conventional autoclaving, gas, or radiation disinfection in hot air ovens, or sterilized
by immersion in decontamination solutions. Remember to remove the rubber
bladder if you use this method
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...