IBP Monitoring
15-12
2. Perform the pressure transducer zeroing beforehand. If the zeroing succeeds,
open the stopcock to the sphygmomanometer.
3. In the IBP PRESSURE CALIBRATE menu, select a value from the CH1 CAL
VALUE or the CH2 CAL VALUE popup menu. The calibration pressure value
of channel 1 or channel 2 is set.
4. Inflate using the inflation orb, until the mercury volume of the
sphygmomanometer rises to the preset calibration pressure value.
5. Adjust the preset calibration value repeatedly until it is the identical to the
pressure value indicated the mercury column.
6. Select CALIBRATE in IBP PRESSURE CALIBRATE menu. The monitor
starts calibrating.
7. Wait for the calibration result. Take actions in response to the prompt
information given by the monitor.
8. After the calibration, disassemble the blood pressure tubing and the T-shape
connector. Then, connect the pressure transducer with the patient, as directed,
for normal monitoring.
NOTE
z
Perform the calibration prior to the use of a new transducer, or at
intervals according to the hospital regulation.
z
The calibration ensures accurate measurement of the monitor. The
pressure transducer zeroing should be conducted before the calibration.
WARNING
z
Never perform the calibration during patient monitoring.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...