IBP Monitoring
15-13
Prompt Information Related to Calibration
Take channel 1 as an example. The monitor may give the following prompt
information after the calibration has been finished.
CH1 SUCCESSFUL CALIBRATE
Channel 1 works normally and can be applied to IBP monitoring.
CH1 SENSOR OFF, FAIL
Verify the transducer of channel 1 is connected to the monitor, and then perform the
calibration again. If the monitor continues to give this prompt information, contact
our Customer Service.
IN DEMO, FAIL
Make sure the monitor is not in DEMO mode, and perform the calibration again. If
the monitor continues to give this prompt information, contact our Customer
Service.
PRESSURE OVER RANGE, FAIL
Verify the preset calibration value is reasonable, and perform the calibration again. If
the prompt information is still displayed, contact our Customer Service.
PULSATILE PRESSURE, FAIL
Verify the pressure value indicated by the sphygmomanometer keeps unchanged,
and then perform the calibration again. If the prompt information is still displayed,
contact our Customer Service.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...