NIBP Monitoring
13-7
13.4 NIBP Setup Menu
Selecting the NIBP label in the parameter area opens the following menu.
Figure 13-2 NIBP Setup Menu
You can perform the following settings in this menu.
ALM
NIBP alarm on/off status
ON: When a NIBP alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a NIBP alarm occurs, the monitor neither
gives alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the
right of the NIBP label.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
ALM REC
Alarm recording
ON: When an NIBP alarm occurs, the monitor enables the
recording;
OFF: When a NIBP alarm occurs, the monitor does not
enable the recording.
SYS ALM HI
Determines the upper limit of the systolic pressure.