Installation and Maintenance
3-5
WARNING
z
Do not use three-wire to two-wire adapter with this instrument.
z
To avoid unexpected power interruption, do no use power outlet with a
wall-mounted switch control.
3.1.5.2 Installing
the
Battery
If the monitor is to be powered by the internal battery, install the battery following
the steps as below:
1. Slide the battery door toward the rear of the monitor to open it.
2. Move the battery catch above to one side using one finger.
3. Insert a battery into the battery slot.
4. Move the battery catch to another side, and then insert the other one in the same
way, if your monitor is equipped with two batteries.
5. Release the battery catch, and it will fix the battery.
6. Close the battery door.
WARNING
z
Make sure the battery door is securely latched. Falling batteries could
seriously or fatally injure a patient.
3.1.5.3 Equipotential Grounding
When other equipments are used together with the monitor, a grounding cable
should be used to connect the equipotential grounding connectors of the monitor and
of other equipments. This helps to reduce the potential differences between different
pieces of equipment, and ensure the safety of the operator and patient.
WARNING
z
If the grounding system is in doubt, the monitor must be supplied from
its internal battery.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...