Appendices
20-39
C.2.16
System Alarm Messages
Alarm message
A
B
L
Cause
Measure
REAL CLOCK
NEED SET
No
No
1
The system time is incorrect.
Reset the system time and then
restart the monitor.
REAL CLOCK
NOT EXIST
No
No
1
No button battery, or the
battery power is depleted.
Add, or replace with a new button
battery.
KEYBOARD INIT
ERR N
No No 1
Note: N represents the error number.
Keyboard error. The keyboard
cannot be used.
KEYBOARD
ERROR
No No 2
NET INIT ERR
(G.)
No No 2
NET INIT ERR
(Ram)
No No 2
NET INIT ERR
(Reg)
No No 2
NET ERR (Run 1)
No
No
2
NET ERR (Run 2)
No
No
2
The system cannot be
connected to the network due
to problems in the monitor’s
network part.
Contact our company for repair.
12V TOO HIGH
No
No
1
12V TOO LOW
No
No
1
A problem occurs to the system
power.
If this alarm message is given for
many times, contact our company
for repair.
BATTERY TOO
LOW
No
No
1
The battery voltage is too low. Connect the monitor with AC
power to recharge the battery.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...