Appendices
20-29
C.2.3 RESP
Module Alarm Messages
Alarm message
A
B
L
Cause
Measure
RESP DISTURBED No No
3
The module circuit is disturbed.
RR EXCEED
No No
1
The circuit is disturbed and the
measurement is inaccurate.
If the problem occurs
continuously, restart the
monitor. If it still exits, contact
our company for repair.
C.2.4 TEMP
Module Alarm Messages
Alarm message
A
B
L
Cause
Measure
TEMP1 SENSOR
OFF
No Yes
3
The T1 sensor is not connected
correctly to the patient or the
monitor.
Check for correct connection of
the T1 sensor.
TEMP2 SENSOR
OFF
No Yes
3
The T2 sensor is not connected
correctly to the patient or the
monitor.
Check for correct connection of
the T2 sensor.
TEMP SELFTEST
ERROR
No No
1
Circuit fault of the temperature
channel.
Contact our company for repair.
TEMP
CALIBRATION
ERR
No No
2
Error in temperature channel
calibration.
Restart the monitor. If the error
remains, contact our company
for repair.
C.2.5 NIBP
Module Alarm Messages
Alarm message
A
B
L
Cause
Measure
NIBP SELFTEST
ERR
Yes Yes 1
Error in NIBP initialization.
Select RESET from the NIBP
SETUP menu. If the error
remains, contact our company
for repair.
LOOSE CUFF
No
Yes 3
The NIBP cuff is not properly
connected.
AIR LEAK
No
Yes 3
PNEUMATIC
LEAK
No Yes 3
Leak in the airway.
CUFF TYPE ERR
No
Yes 2
The cuff applied is not
appropriate to the patient type.
AIR PRESSURE
ERROR
No Yes 3
WEAK SIGNAL
No
Yes 3
Failure occurs in the pulse
measurement. The monitor cannot
perform measurement, analysis,
Check the patient’s condition,
and check if the patient type is
correct. Replace with a proper
cuff and connect it correctly. If
the problem still exists, contact
our company for repair.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...