ECG/RESP Monitoring
11-40
ALM REC
ST alarm recording
ON: The monitor starts recording when an ST alarm occurs;
OFF: The monitor does not record when an ST alarm occurs.
ANT ALM HI
Determines the upper limit of the ST alarm in the ANT group;
2.0mV is the highest.
ANT ALM LO Determines the lower limit of the ST alarm in the ANT group;
-2.0 mV is the lowest.
INF ALM HI
Determines the upper limit of the ST alarm in the INF group;
2.0mV is the highest.
INF ALM LO
Determines the lower limit of the ST alarm in the INF group;
-2.0 mV is the lowest.
LAT ALM HI
Determines the upper limit of the ST alarm in the LAT group;
2.0mV is the highest.
LAT ALM LO
Determines the lower limit of the ST alarm in the LAT group;
-2.0 mV is the lowest.
12-lead review
Select 12 LEAD RECALL >> In ECG SETUP menu. The following menu appears.
Figure 11-28 12-lead review
In this menu, you can recall 80 12-lead analysis results. When the number of 12-lead
results exceeds 80, the earliest results will be deleted.