ECG/RESP Monitoring
11-37
GAIN
Used to adjust the amplitudes of ECG waveforms for all leads.
Option:
×
0.125,
×
0.25,
×
0.5,
×
1,
×
2, AUTO
If GAIN is set to AUTO, the gain will be adjusted by the monitor
automatically. Each waveform is followed by a 1mV scaling line,
the height of which is proportional to the amplitude of the
waveform.
RHYTHM
Used to select the name of the rhythm lead. Option: I, II, III,
aVR, aVL, aVF, V1, V2, V3, V4, V5 and V6.
CATCH
When the waveform is long enough, you can select this label to
freeze all waveforms except the rhythm lead waveform, and this
label changes to RELEASE; Press it again, the frozen waveforms
will be unfrozen, and this label becomes CATCH again.
When the waveforms are not long enough, if you select this label,
a dialog box will appear prompting that the analysis cannot be
done due to the inadequate data.
RECALL
When the waveforms are frozen, you can select this label and
rotate the control knob clockwise or counterclockwise to review
the waveforms.
ANALYSIS
When the waveforms are frozen, you can select this label to
perform the 12-lead analysis. During the analysis, the monitor
does not respond to any operation. At the end of the analysis, the
analysis result will be displayed on the screen.
Figure 11-25 ECG 12-lead analysis result
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...