CO
2
Monitoring
17-17
Other Setup
Selecting the OTHER SETUP option opens the following menu.
Figure 17-9 CO
2
Other Setup Menu
You can perform the following settings in this menu.
MAX HOLD
Maximum holding time
It determines the maximum holding time of the CO
2
parameters.
Options: OFF, 10S, 20S and 30S.
BTPS
Body temperature and pressure, Saturated
Options: ON and OFF.
When BTPS is turned ON, the system performs BTPS compensation automatically;
when BTPS is turned OFF, the BTPS compensation is disabled.
When measuring the CO
2
content inside the patient’s lung, turn ON BTPS. Normally,
the lung temperature of the patient is 37
℃
and the water vapor is sufficient or
regarded as saturated. But the temperature and water vapor content of the sampling
line are different from the patient’s exhalation. Thus, when the patient’s exhalation
passes the sampling line, the accuracy of the CO
2
concentration measured by the
sensor will be adversely affected. To ensure an accurate measurement, you should
turn ON the BTPS. In this mode, the system adjusts the calculation coefficient
automatically, so that the measured value is just the actual CO
2
content in the
patient’s lung.
When measuring the CO
2
content of the environment or a certain vessel, turn OFF
BTPS. In the standard ambient pressure and dry gas, the accuracy of the sensor
measurement is not affected.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...