WARNING
z
Setting the SpO
2
upper alarm limit to 100% will disable the upper alarm
limit. High oxygen levels may predispose a premature infant to
retrolental fibroplasia. Therefore, the upper alarm limit for oxygen
saturation must be carefully selected in accordance with the commonly
accepted clinical practices.
SWEEP
Waveform speed
Options: 12.5 and 25.0 mm/s.
PR SOUND
Pulse volume
Range: 0
–
10. 0 indicates the volume is closed and 10
indicates the maximum volume.
SENSITIVE
Sensitivity of SpO
2
calculation
Options: HIGH, MED and LOW.
When HIGH is selected, the monitor gives quick response to
the change of the oxygen saturation but the measurement
accuracy might be compromised.
When LOW is selected, the monitor gives slow response to
the change of the oxygen saturation but the accuracy is
relatively high.
NIBP SIMUL
ON: When measuring NIBP and SpO
2
at the same side, the
SpO
2
alarm status remains unchanged until the NIBP
measurement is finished;
OFF: The function of NIBP SIMUL is disabled.
DEFAULT
You can select this option to access the SPO2 DEFAULT
CONFIG menu, in which you may select FACTORY
DEFAULT CONFIG or USER DEFAULT CONFIG. After
finishing your selection and exiting the menu, the system
pops up a dialog box asking for your confirmation.
Summary of Contents for PM-9000
Page 1: ...PM 9000 Patient Monitor Operation Manual...
Page 2: ......
Page 14: ...Contents XII FOR YOUR NOTES...
Page 42: ...The Basics 2 18 FOR YOUR NOTES...
Page 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Page 106: ...Alarms 6 12 FOR YOUR NOTES...
Page 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Page 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Page 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Page 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Page 377: ......
Page 378: ...P N 9100 20 11311 6 2...