170 Keysight N9923-90001 User’s Guide
VVM (Vector Voltmeter) Mode - Option 308
1-Port Cable Trimming Measurements
6.
Attach the shortest cable to the FieldFox
PORT 1
connector. This cable is
the reference cable. Leave the end of the reference cable open
(unterminated).
7.
Press
Meas Setup 4
then press
Zero
.
8.
Remove the reference cable from the FieldFox
PORT 1
connector.
9.
Connect an unterminated, longer cable to the FieldFox
PORT 1
connector.
10.
Observe the displayed phase difference between the reference cable and
the untrimmed cable.
11.
Carefully trim the cable until the phase shift reads zero. The attached
cable’s electrical length is now matched to the reference cable.
12.
Repeat steps 10 through 12 for the remaining cables to be trimmed.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...