Keysight N9923-90001 User’s Guide 169
VVM (Vector Voltmeter) Mode - Option 308
1-Port Cable Trimming Measurements
1-Port Cable Trimming Measurements
A 1-Port Cable Trimming measurement is used in a cable fabrication procedure
to validate proper electrical length.
How to make a 1-Port Cable Trimming Measurement
1.
Press
Mode
then
VVM
then
1-Port Cable Trimming
.
2.
Press
Freq/Dist
then enter the frequency for the measurement.
3.
Press
Meas Setup 4
then
Zero OFF
.
4.
Optionally calibrate the FieldFox. Learn more about VVM calibration in
.
—
If using an adapter to connect the cables to the FieldFox, press Cal
5 then select either QuickCal or Mechanical Cal. Follow the FieldFox
calibration prompts. During calibration, the measurement
(S-parameter) is referred to as S11.
—
If connecting the cables directly to the FieldFox test port, CalReady
is very accurate.
5.
To ensure that the calibration is adequate, perform the following
optional
Measurement Verification
procedure.
a.
Connect a short standard to
PORT 1
where cables to be trimmed
will be connected.
b.
Press
Meas Setup 4
and press
Zero
.
c.
Remove the short standard and connect the open standard.
d.
Confirm that magnitude measurement is less than –0.1 dB and that
the phase value reads 180° degrees.
— The R receiver measures incident
signal (green line) out of the source.
— The A receiver measures reflected
signal (red line) from the end of the
DUT (black line).
— In VVM mode, the display ALWAYS
shows ratioed (A/R) measurements
which is the electrical length of the
DUT.
This measurement verification procedure is suitable for low frequency
measurements. However, for high frequency measurements, a more
sophisticated procedure is required.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...