176 Keysight N9923-90001 User’s Guide
Data Analysis Features
All about Markers
How to Create a Delta Marker
— Create a
Normal
marker and move it to the reference position.
— Press
Delta
. The marker becomes the reference marker and fixed in its
X-Axis position.
— Move the delta marker using the numeric keypad, the
arrows, or the
rotary knob.
— Press
Enter
when finished.
How to move the Reference marker
— Press
Delta
to turn the Delta marker OFF.
— Move the marker to the new position.
— Then press
Delta
again, to turn the Delta marker back ON.
Marker Table
The Marker Table can be displayed at the bottom of the FieldFox screen. It can
display information for up to 6 markers in a full-width window, and up to 3
markers for a half-width window (NA mode). The marker table “squeezes” the
graticule area when activated.
How to display the Marker table
— Press
Marker
.
— Then
More
— Then
Marker Table On Off
.
Coupled Markers
Coupled Markers, available in NA mode only, allows you to simultaneously
create and move markers that have the same number for all traces. For
example, when marker 2 is created on a trace, a marker 2 is created for ALL
traces. When any marker 2 is moved, they all move together to the same
X-axis position.
— Coupled markers are allowed only in NA mode and are relevant only in
multi-trace configurations.
— You can have markers that are coupled and others that are uncoupled.
— Peak functions move the marker to the peak on the
active trace
, while all
other markers move to the same X-axis position.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...