168 Keysight N9923-90001 User’s Guide
VVM (Vector Voltmeter) Mode - Option 308
VVM Mode Settings
Press
Zero OFF
before performing a QuickCal or Mechanical Cal (sets Zero to
OFF).
How to select Zeroing
— Press
Meas Setup 4
— Then
Zero Ref
—
Open
– select to display Phase = 0.0 Deg when zeroed.
—
Short
- select to display Phase = 180.0 Deg when zeroed.
— Then
Zero ON
— Select
Zero OFF
to turn zeroing off.