80 Keysight N9923-90001 User’s Guide
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
—
Ref Level
Manually set the value of the reference line.
Enter a negative value by pressing
Run/Hold (+/-)
either
before or after typing a value
—
Ref Position
Manually set the position of the reference
line. Values must be between
0
(TOP line) and
10
(BOTTOM line)
3. Set Top and Bottom graticule values. The scale per division is
calculated.
—
Top
to set the value of the Top graticule.
—
Bottom
to set the value of the Bottom graticule.
—
Enter a negative value by pressing
Run/Hold (+/-)
either
before or after typing a value.
Averaging
Trace Averaging helps to smooth a trace to reduce the effects of random noise
on a measurement. The FieldFox computes each data point based on the
average of the same data point over several consecutive sweeps.
Average Count determines the number of sweeps to average; the higher the
average count, the greater the amount of noise reduction.
An average counter is shown in the left edge of the screen as
Avg N
. This
shows the number of previous sweeps that have been averaged together to
form the current trace. When the counter reaches the specified count, then a
‘running average’ of the last
N
sweeps is displayed. Average Count = 1 means
there is NO averaging.
This setting can be changed at any time without affecting calibration accuracy.
How to set Trace Averaging
— Press
BW 2
.
Scale annotation on the FieldFox screen
Reference Line = red arrow
Ref Level = -40 dB
Ref Position = 1
Scale = 2 dB per division
Averaging is often used to increase the dynamic range of a measurement.
To achieve the highest dynamic range, select NA mode and reduce the IF
Bandwidth setting. Learn more about dynamic range in
.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...