172 Keysight N9923-90001 User’s Guide
VVM (Vector Voltmeter) Mode - Option 308
A/B and B/A Measurements
A/B and B/A Measurements
These measurements allow you to use the FieldFox A and B receivers, along
with an external source, to make ratioed measurements.
The image below shows a typical configuration for a B/A magnitude and phase
measurement of a 2-port DUT. This DUT can be an antenna / probe setup.
Using network analyzer terms, the image shows the A receiver as the reference
receiver which measures the signal that is applied to the DUT. The B receiver is
the test port receiver which measures the DUT response.
Figure 11-4
B/A Transmission measurement with external source
Procedure
1.
Connect the external source’s frequency reference output to the FieldFox
Freq Ref IN
on the top panel. If you are unable to use a common frequency
reference, then increase the IFBW to ensure you are measuring the peak of
the correct signal. After locating and tuning to the signal, you may be able
to lower the IFBW to reduce noise.
2.
On the FieldFox, select
System 7
, then
System Configuration
, then
Frequency Ref
, then
Freq Ref Source
then
Ext
.
3.
Press
Mode
, then
VVM
, then
B/A
or
A/B
4.
For highest accuracy, set the FieldFox output power to Low. Press
Meas
Setup 4
, then
More
, then
Output Power Low
.
5.
To calibrate, replace the DUT with a THRU connection, preferably by
connecting together the two Power Splitters (shown in green in the
graphic above). For low frequency measurements, TEEs can be used.
6.
Press
Meas Setup 4
, then
Zero
.
7.
Connect the DUT and measure.
These measurements require the “Full 2-port S-parameters” option.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...