68 Keysight N9923-90001 User’s Guide
Time Domain – Option 010
Time Domain (Transform) Settings
How to select Window settings
— Press
Meas Setup 4
.
— Then
Settings
.
— Then press
arrows to move to the Window row.
— Then press
Edit
.
— Then press
Window
and choose from the following:
—
Maximum
(Maximum – Optimized for dynamic range)
the noise floor
is lowered to provide the ability to measure low-level responses.
(Default setting)
—
Medium
– Compromise between Min and Max window settings.
—
Minimum
– Best Response Resolution, providing the ability to resolve
between two closely-spaced responses.
— Then press
Done Edit
.
— Again press
Done
.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...