216 Keysight N9923-90001 User’s Guide
System Settings
Preferences
—
Espanol – Spanish
—
Deutsch – German
—
Italiano – Italian
—
Francais – French
—
Russian
—
Japanese
—
Chinese
—
Turkish
Battery Saver
When the battery saver is OFF, the RF source power is left ON at the end of the
sweep which consumes more battery power. This feature is used in CAT, NA,
and VVM modes when the internal source is turned ON.
Battery Saver is the ONLY FieldFox setting that must be made as a Preference
from the Quick Settings table. This setting remains until you change it again.
How to set Battery Saver
— Press
System 7
— Then
Preferences
— Then
Preferences
— Then scroll to
Global Settings
>
Battery Saver
—
ON
– To conserve battery power, the RF source power is turned OFF at
the end of each CAT, NA, and VVM mode sweep. This behavior does
not impact FieldFox measurements.
—
OFF
– Source power is kept ON at all times, including during trigger
Hold mode and between single sweep acquisitions. The RF source
stays at the stop frequency until another sweep begins. This provides
a constant, uninterrupted RF output from the RF OUT port. This
behavior also provides the highest trace stability sweep-to-sweep
which may only be evident while using either single sweep or
Run/Hold. With Battery Saver OFF, the use of a full battery charge is
reduced by approximately 45 minutes.
The FieldFox internal temps will run 2°C – 5°C cooler with the Battery
Saver on (the temperature range is dependent on the ambient temp).
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...