File Management
Saving and Recalling Files
—
USB
Saves/Recalls/Moves (Copies) files to or from a USB Flash drive
(not included) that is inserted in one of the USB connectors. Only the
root folder is available.
How to set File Type
— Press
Save/Recall 9
— Then
File Type
. Choose one of the following:
—
State
Saves ALL FieldFox instrument settings for ALL Modes to a *.sta
file. If a calibration is ON in CAT or NA Mode, calibration data is also
saved. This file can be opened ONLY by a FieldFox. State files can be
saved between compatible FieldFox models. See note in.
.
—
Trace+State
Saves the current trace and instrument settings to a *.sta
file. In SA Mode, all traces are saved. This file can be opened ONLY by
a FieldFox. When recalled, the FieldFox sweep is set to Hold.
—
Picture
Saves a 640 x 453 pixel image of the FieldFox screen to a *.png
file. The picture does NOT include the softkeys.
—
Data (csv)
Saves trace data from the current Mode to a *.csv file. This
format can be imported into spreadsheet software such as Microsoft
Excel. Column A is always the frequency at which the data is
measured.
——In
CAT
and
NA
Modes, only the data trace is saved – not the
memory trace.
——In
Power Meter
and
VVM
Mode, NO data is saved.
—
Data (S1P) / (S2P)
Saves
CAT
and
NA
Mode trace data to an *.S1P or
*.S2P file, depending on the active measurement. This file format is
used by CAE programs such as Keysight's Microwave Design System
(MDS) and Advanced Design System (ADS). It can also be imported
into spreadsheet software such as Microsoft Excel.
About SNP Files
The SNP file includes header information, and the following data:
— S1P - saves 1-port (receiver) data for the active data trace. *.s1p Files
contain 1 stimulus value and 1 S-parameter (total of 3 scalar values)
Stim
Real(Sxx) Imag(Sxx)
— S2P (NA Mode ONLY) - saves 2-port data for the active trace. *.s2p
Files
contain 1 stimulus value and 4 S-parameters (total of 9 scalar values)
Stim
Real(S11) Imag(S11) Real(S21) Imag(S21) Real(S12) Imag(S12)
Real(S22) Imag(S22)
Insert only ONE USB Flash Drive in the FieldFox when saving or copying
files.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...