132 Keysight N9923-90001 User’s Guide
Power Meter (USB) Mode
USB Power Meter Settings
Resolution
Resolution sets the number of digits to display after the decimal point on the
digital power meter display.
How to set Resolution
— Press
Scale / Amptd.
— Then
More
.
— Then, in
Resolution
, choose from 0, 1, 2, or 3.
Averaging
Averaging is used to improve measurement accuracy in low power or noisy
power situations. Increased averaging improves measurement accuracy, but
also reduces measurement speed.
How to set Power Meter Averaging
— Press
BW 2
or
Meas Setup 4
.
— Then choose from the following:
—
Averaging Auto Man Off
—
Auto
Allows the USB Power Sensor to use its Auto Averaging
settings. See the USB Power Sensor documentation for details.
—
Man
Manually enter an averaging count by setting
Num Averages
.
—
OFF
Performs NO averaging.
—
Num Averages
Sets the number of Power Meter readings to be
averaged before a valid reading is displayed. When the number is
changed,
Averaging
is automatically set to Man.
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
Single
button is pressed. Use
Single
to conserve battery
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
— Press
Sweep 3
.
— Then choose one of the following:
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...