Keysight N9923-90001 User’s Guide
83
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
Output Power
Set the power level out of the FieldFox to High, Low, or manually set power
level to a value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss
devices to place the signal farther from the noise floor. However, for devices
that are sensitive to high power levels such as amplifiers, use the Low power
setting.
For best measurement accuracy, use the Manual power setting at -15 dBm.
After calibration, the power level can be decreased for amplifiers, or increased
for higher dynamic range.
How to set Output Power
— Press
Meas Setup 4
.
— Then
Power
— Then
Output Power
—
High
(Default setting) Sets output power to the maximum achievable
power at all displayed frequencies. Output power is NOT FLAT across
the displayed FieldFox frequency span. Please see
“Specifications/Data Sheet” on page 263
—
Low
Sets output power to approximately –45 dBm, FLAT across the
displayed FieldFox frequency span.
—
Man
Set output power to an arbitrary value, FLAT across the displayed
FieldFox frequency span. If flattened power can NOT be achieved, a
warning message and beep occurs. To achieve a flattened output
power, reduce the power level or stop frequency.
—Then press
Nominal Power
—Then enter a value using the numeric keypad, the
arrows, or the
rotary knob.
Power Level settings in this mode will NOT change Power Level settings in
other modes. To help prevent damage to your DUT, use caution when
changing modes with your DUT connected to the FieldFox test ports.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...