Keysight N9923-90001 User’s Guide
53
NA (Network Analyzer) Mode
NA Mode Settings
— Then choose from the following:
101
|
201
|
401
|
601
|
801
|
1001
|
1601
|
4001
|
10001
.
— Using SCPI, Resolution can be set to ANY number of points between 3 and
10001. See the Programming Guide at
http://na.support.keysight.com/fieldfox/help/Programming/webhelp/FFP
Sweep Time
The fastest possible sweep time is always used as the default setting. Use the
Min Swp Time setting to slow the sweep time when measuring long lengths of
cable. Learn more at the FieldFox Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
The actual sweep time is shown on the FieldFox screen. See the Screen Tour
on
. To increase the sweep time, enter a value that is
higher than the actual sweep time. The increase will not be exactly the amount
that you enter, as the actual sweep time is the composite of many factors.
How to set Sweep Time
— Press
Sweep 3
.
— Then
Min Swp Time
.
— Enter a value using the numeric keypad.
— Press a multiplier key. Learn about multiplier abbreviations in
.
Output Power
Set the power level out of the FieldFox to High, Low, or manually set power
level to a value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss
devices to place the signal farther from the noise floor. However, for devices
that are sensitive to high power levels such as amplifiers, use the Low power
setting.
For best measurement accuracy, use the Manual power setting at -15 dBm.
After calibration, the power level can be decreased for amplifiers, or increased
for higher dynamic range.
Measurement speed specifications do NOT apply in Temperature Control
Mode. Learn more in
“Temperature Control Mode” on page 25
Power Level settings in this mode will NOT change Power Level settings in
other modes. To help prevent damage to your DUT, use caution when
changing modes with your DUT connected to the FieldFox test ports.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...