186 Keysight N9923-90001 User’s Guide
Data Analysis Features
All about Limit Lines
——
Upper
- FAILS any measured data point
ABOVE
the line.
——
Lower
- FAILS any measured data point
BELOW
the line.
—
Off/On
– Press
Edit
to toggle between the following:
——
On
- Limit line is drawn on the screen and PASS/FAIL testing
occurs.
——
Off
- Limit line is NOT drawn on the screen and testing does
NOT occur.
—
Frequency
– Enter a numeric X-axis value of the data point.
—
Amplitude
– Enter a numeric Y-axis value of the data point.
— Press
Add
then choose from the following:
—
Add Point
- Adds a blank X/Y data point below the cursor.
—
Add Limit
- Adds a default limit line below the current limit.
— Press
Clear
then choose from the following:
—
Clear Point
- Removes the X/Y data point on which the cursor resides.
—
Clear Limit
- Removes the limit Line on which the cursor resides.
—
Clear All
- Removes ALL limit lines.
— Press
<Back
to return to the Limit menu.
Relative Limit Lines
Relative limit lines are X/Y coordinates that are referenced to the center
frequency (X-axis) and Reference level (Y-axis).
Use relative limits to test devices or signals that are identical except for the
center frequency or amplitude. With relative limit lines in place, you change the
center frequency of the measurement and the limits look as though they are in
the same location.
The same is true for Reference Level. When you expect to measure a different
power level, instead of changing the Y-axis value of all of the fixed limits,
simply change to relative limits. The limits will then remain a constant Y-axis
distance from the Reference Level.
Each limit line can be changed at any time from Fixed to Relative, and Relative
to Fixed.
How to Create a Relative Limit Line
The easiest way to create a Relative limit line is to create a Fixed Limit at a
specific center frequency and reference level. Learn how
. Then change the limit line to Relative. The X/Y data points will be
re-calculated relative to the center frequency and reference level.
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...